ASLi (CHINA) TEST EQUIPMENT CO., LTD

AI SI Li (CHINA) TEST EQUIPMENT CO., LTD High Quality Assurance , Good After-sales Coexistence !

Manufacturer from China
Active Member
10 Years
Home / Products / Walk In Stability Chamber /

Pharmaceutical Study Walk In Stability Chamber For Temperature And Humidity Measurement

Contact Now
ASLi (CHINA) TEST EQUIPMENT CO., LTD
Country/Region:china
Contact Person:MsDoris
Contact Now

Pharmaceutical Study Walk In Stability Chamber For Temperature And Humidity Measurement

Ask Latest Price
Video Channel
Brand Name :ASLi
Model Number :THR-34500-A
Certification :CE,SGS, 3rd party calibration
Place of Origin :China,DongGuan
MOQ :1
Price :Negotiation
Payment Terms :T/T, Western Union
Supply Ability :30 sets per month
Delivery Time :40~45 workdays
Packaging Details :Standard export polywood case packaging
Interior Dimension :W5000 X H2300 x D3000 mm
Exterior Dimension :W5900×H2800×D3850 mm
Test Points :30 Deg C & 65%R.H.
Temperature range :25~60 Deg C
Humdity Range :40%~90%R.H.
Power :AC380V, 3 phase 5 lines
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

30 Deg C & 65%R.H. Pharmaceutical Study Walk In Stability Chamber CE Certification

Model: THR–34500-A

Dimensions:

Interior Dimension: W5000×H2300×D3000 mm

Exterior Dimension: W5900×H2800×D3850 mm

Tech Spec:

Temperature :25℃~60℃

Temperature Indication Resolution: ±0.1℃

Temperature Control Accuracy: ±0.5℃

Temperature distribution uniformity:±3 ℃

Humidity: 40%~90%R.H

Humidity Indication Resolution: ±0.1%R.H

Humidity distribution uniformity: ± 5.0%R.H

Test Points: 30 ℃ & 65%R.H.

Measuring and Control System:

1 High-accuracy platinum resistance is used for temperature and humidity measurement.

2 Central Control Unit:

2-1 LCD touch screen controller;

2-2 Machine conversational mode is used for control system;

2-3 Two lines of input & two lines of output control mode for controlling temperature and humidity separately; Can

display set point, program curve, working time, and working condition of heaters, refrigeration units & fan.

2-4 Test procedure automatic operation and PID parameter self-tuning function;

2-5 Can automatically combine the subsystem like heating system, humidification/dehumidification system,

refrigeration system, circulation fan, super temperature alarming to operate, to ensure high quality of the whole temperature/humidity control system;

2-6 Control system has automatic self-checking, self-diagnosis, fault display and alarming function.

Pharmaceutical Study Walk In Stability Chamber For Temperature And Humidity Measurement

Main Parameters:

1 Display Mode: LCD display, touch control

2 Display Accuracy: temperature 0.1%, huidity0.1%

3 Set Accuracy: temperautre0.1℃, huidity0.1%

4 Display Resolution: temperautre0.1℃, humidity0.1%

5 Sampling Time: 72 S/Time;

6 Graphical Display: Full display of setting and program curve

7 Program Capacity: Max.269 programs,each program with 50 Segment, max.13450 cycles

8 Temperature/humidity PV value transfer: this instrument has two way 4-20mA analog output,

Synchronous transmission temperature and humidity; Multi Power Failure Recovery function

9 Time display in /Y/M/D/ mode, also the operating time and program running time can be displayed

10 Communication: USB/RS-232 communication interface, dedicated software are provided.

11 Main Accessories: "LG" circuit breakers, leakage protector, AC contactor, thermal relay device;

12 System Running Mode: program operation, constant operation

13 Remote Control and Monitoring: this machine is equipped with remote communication software,

communication interface and communication cable for remote control & monitoring. Data record, collecting curve , data acquisition, date save, printing can be finished.

14 Humidification/Water Supply Systems:It is mainly composed of built-in water tank, wet-bulb, solenoid

valve, humidifier, water level controller, connecting pipe, etc.

15 Humidification water processor (pure water treatment device, optional), can automatically/manually

Adding water; soft water after 3 levels filtration, the water resistivity is greater than 500 Ω m.

Pharmaceutical Study Walk In Stability Chamber For Temperature And Humidity MeasurementPharmaceutical Study Walk In Stability Chamber For Temperature And Humidity MeasurementPharmaceutical Study Walk In Stability Chamber For Temperature And Humidity Measurement

Multi protection function:

Grounded protection

Leakage/breakage protection

Heater short-circuit protection

Blower motor overload protection

Refrigerator over-voltage protection

Refrigerator overload protection

Water shortage protection

Using Conditions:

Environment Temperature: 25℃±5℃;

Environment Humidity: < 85%R.H. (25℃);

Air Pressure: 86 to 106Kpa;

Inquiry Cart 0