Add to Cart
Standard Parameters
Model | THSH-2000-B |
Internal Dimension | 2000×1000×1000(W×D×H) mm |
External Dimension | 3000×1500×1950(W×D×H)mm |
Cabinet capacity | 2000L |
Controlled parameters | Temperature , relative humidity, Salt fog |
Drying temp. range | Ambient to +75 deg C |
Humidity temp. range | Ambient to +60 deg C |
Humidity range | 15% to 95% R.H adjustable(Temp.60deg C ~ 75deg C) |
Salt fog fallout rates | rates adjustable from 0.5 to 2.5ml per 80cm2 per hour |
Chamber material | P.V.C rigid plastic boar |
Spraying method | Continuous and Programmable spray type |
Temperature Rising/Falling Speed Rate: | From R.T.( 25 Deg C) ~ 75 Deg C, needs about 30 mins, ( approx. 2.0 Deg C /min) |
Accuracy | Temperature 0.1 Deg C ; Humidity 0.1 %; |
Control Accuracy | Temperature ±0.5 Deg C; Humidity ± 2.5%R.H. |
Uniformity | Temperature: ± 3.5 Deg C |
Multiple Safety Protection Devices | Current discharge protection, over pressure protection, over |
4.Mechanical Structure
1. The internal chamber & testing made of CPVC sheet ; |
2. Salt water adding barrel with hidden water level observation gauge, easy to clean, and not easy broken. |
3. Saturated air bucket with SUS#304 stainless steel, pressure resistant barrel to ensure good heat preservation effect. |
4. Testing room sample shelf adopts plane graduated rack, free adjustment to the angle, fog and mist completely consistent of the four corners, more quantity of test pieces placed. |
5. Salt Spray Temperature (Able to set constant temperature control)
(A)Saline Test : NSS,ACSS | 1. Testing Chamber : 35 Deg C (+/-)2 Deg C |
2. Saturated Air Bucket: 47 Deg C (+/-)2 Deg C | |
(B) Corrosion Test: CASS | 1. Testing Chamber Temperature: 50 Deg C (+/-) 2 Deg C |
2. Saturated Air Bucket: 63 Deg C (+/-)2 Deg C |
6. Air Supply System
1.Air pressure is 1 kg/cm2, divided into two sections adjustment |
2.First step is rough adjustment: 2 kg/cm2 imported type air filter with drainage function, the second step is precision adjustment: 1kg/cm2 Pressure gauge |
7. Spray Method
1. Absorb brine to atomization, uniform atomization degree, to ensure continuous testing standard. |
2. Precise Glass Spray Nozzle: made of special glass,able to adjust spray volume and spray angle, ensure without salt crystal blocking working withing 5000 hours and fog uniformly. |
3. Fog Spray quantity: adjustable type 1-2ml(ml/80cm2/h) (average quantity of 16 hours) |
ASLi main products:
Environmental chambers, temperature (humidity) chamber, PCT /Hast chamber (Highly accelerated stress test chamber),ESS chamber, Walk in stability chamber,Vacuum oven,vibration tester, mechanical shock tester,environmental combined vibration chamber,etc
Package information:
We have professional packing workers to pack your products!
1.Out packing:Standard export wooden case;
2.Inner packing:With careful stretch film+ hard cardboard;