Add to Cart
PLC Control Touch Screen Temperature Humidity Electrodynamic Vibration Shaker Chamber
1.Model: THV -1331-F
2 Application
Vibration chambers combined with temperature and/or humidity are available for temperature & vibration testing or temperature, humidity and vibration.Our this combined test system perform tailored environmental stress screening & reliability tests to reduce product development time,reveal hidden manufacturing defects and improve product reliability.
3-1 The Whole Temperature Humidity Chamber for Electrodynamic Shaker
3-1-1 Dimension
(1) Interior : W1100 × H1100× D1100 mm
(2) Exterior : W3200 × H2600× D2500 mm
3-1-2 Material
(1)Interior is SUS#304 stainless steel
(2) Exterior is SUS#304 stainless steel or paint coated
(3) Thermal insulation materials: High density fire-resistant PU foam insulation material
3-1-3 Mechanical Structure
(1)Single door with a viewing window ( 450×300×40mm, three vacuum layers) on the front chamber
(2)Electric cabinet, refrigerator and air conditioner are in back of electric cabinet
(3)The dehumidifier is in the back of the chamber.
3-1-8 Control System
(1)Programmable LCD touch screen controller, viewing angle 256x128 ,with back light
(2)Temperature ramp control : 0.1 ~ 9.9
(3)Upper and lower limitation standby and alarm function
(4)Temperature and humidity sensor is provided
(5)6 groups P.I.D control parameter setting & P.I.D auto graphical function
(6)Both set point and actual value of temperature/humidity can be directly displayed
(7)Operating time adding up function
(8) Backup function in case of power off
(9)RS-232 communication
(10)Real time graphical curve display function while programmable running
(11)Can book in advance to start and shut down
(12)Date and time can be adjusted
(13)Screen lock function
3-2 Technical Parameters for Temperature Humidity Chamber for Electrodynamic Shaker
(1) Temperature Range:D: -70 ℃ to 150℃ (adjustable& controllable),
(2) Humidity Range: 20%~95%R.H. (adjustable& controllable)
(3) Temperature Rising Velocity: Approx. 0.5~3 ℃/min; (No load with no live control)
(4) Temperature Falling Velocity: Approx. 0.5~1.0℃/min; (No load with no live control)
(5) Indication Resolution: Temperature: ±0.1℃; Humdity:±0.1%R.H.
(6) Control Accuracy: Temperature ±0.5℃; Humdity:±2.0%R.H
(7) Distribution Uniformity:
Temperature : -70℃~100℃, ±2.0℃; temperature 100℃~150℃, ±3.0℃
Humdity: ≤80%R.H. ±5.0%R.H; Humidity≥80%R.H. ±3.0%R.H